| INPUT |
240 Volts AC 50 Hz singIe Phase. |
| OUTPUT
|
0 - 30 KV DC with one end at ground potential.
|
| VOLTAGE SOURCE |
Double wound H.T. Transformer and rectifiers. |
| RATING
|
15minutes ON 15minutes OFF. |
| METERING
|
0-15-30 kV and 0-15-30 mA. |
| |
Accuracy : ± 4% of full scale deflection. |
| PROTECTION
|
Fuse, a fast acting dc relay isolates the H.T.
when H.T. current exceeds the set trip value. |
| INTERLOCKING
|
Zero - start interlocking for H.V. output. |
| AUXILIARIES
|
Mains switch, fuses, indicating lamps for
MAINS ON, HT.ON, AND H.T.OFF. Push-buttons for
H.T. ON and H.T. OFF, contactor operated. |
| TERMINATION |
H.T. End on suitable Insulator and other end
on earth potential. |
| ENCLOSURES |
The unit will be in two parts: |
| |
Control unit : Houses switches, fuse,
lamps, variac etc in a sheet metal case. |
| |
HT. Transformer Unit : H.T. transformer
and rectifers. |
| OPTIONS |
| : Digital meters |
: Motorized rise of voltage |
| : Multi range meters |
: Audio Alarm |
| : Any other feature as required |
|
|