| INPUT |
240 Volts AC 50 Hz singIe Phase. |
| OUTPUT
|
0 - 50 KV RMS AC 50 Hz. with one end at ground
potential. |
| VOLTAGE SOURCE |
Double wound oil cooled H.T. Transformer |
| RATING |
15minutes ON 15minutes OFF |
| METERING |
KV 0-25-50 and mA 0-50-100. |
| |
Accuracy : ± 4% of full scale deflection. |
| PROTECTION |
Fuse, a fast acting dc relay isolates the H.T.
when H.T. current exceeds the set trip value. |
| INTERLOCKING |
Zero - start interlocking for H.V. output. |
| AUXILIARIES |
Mains
switch, fuses, indicating lamps for MAINS ON, HT.ON, AND H.T.OFF.
Push-buttons for
H.T. ON and H.T. OFF, contactor operated. |
| TERMINATION |
H.T. End on suitable Insulator and other end on
earth potential. |
| ENCLOSURES |
The unit will be in two parts: |
| |
Control unit: Houses switches, fuse,
lamps, variac etc in a sheet metal case. |
| |
HT. Transformer Unit: H.T. transformer
oil cooled with unidirectional rollers. |
| OPTIONS |
| :
Digital meters |
: Motorized rise of voltage |
| : Multi range meters |
: Audio Alarm |
| : Any other feature as required |
|
|