| INPUT |
240 Volts AC, ± 10% , 50 / 60 Hz. |
| OUTPUT |
A) CURRENT OUTPUT: |
| |
1) 0 - 50 Ampere @ 20 Volt |
| |
2) 0 - 25 Ampere @ 40 Volt |
| |
3) 0 - 2.5 Ampere @ 150 Volt |
| |
4) 0 - 1 Ampere @ 250 Volt |
| |
Power : 1000 VA. |
| |
Each of these outputs has a separate terminal referred to
one common terminal. |
| |
B) VOLTAGE OUTPUT : |
| |
1) 0-240 Volt DC/ 1Amp continuously adjustable. (This
is not a floating output) |
| |
2) 0-240 Volt AC/ 1Amp, continuously adjustable. (This
is fully floating output) |
| |
Each of these outputs has a separate pair of terminals. |
| DUTY CYCLE |
20 minutes ON/20 minutes OF |
| METERING |
Current / Voltage : 3 1/2 Digit 12 mm LED. |
| |
Accuracy : ± 1% of rdg. ± 3 digits. |
| |
a) 0-2A a c ( for output 0-1A/250V ac; resolution
0.001 A)
b) 0-50A a c ( for output 0-2.5A, 0-25A and 0-50A ac; resolution
0.1 A)
c) 0-250V d c
d) 0-250V a c. |
| |
2) Voltage : Reads outputs B in volts. Accuracy : 0.5% of
rdg. ± 3digits. |
| TIMER |
4 Digit, 12mm LED, with crystal time base. |
| |
Auto ranging from 0.0001 to 9999 seconds. Accuracy: ± 0.3%. |
| TIMER CONTROL CIRCUITS |
The timer automatically starts when current output
is initiated and stops automatically when the device under test
operates. |
| |
Test Conditions : |
| |
Contact : NC / NO Contact. |
| AUTO-CUT-OFF |
Cuts off current output on operation of item. |
| ACCESSORIES |
(a) Operating Manual with circuit diagram. |
| |
(b) 3 Core 2 meter long Main Cord |
| |
(c) 3 meter long 50 Amp test lead pair. |
| |
(d) 3 meter long Voltage AC / DC test leads pair. |